image
image
Fig.1:Pumping speed vs. gas load illustration
Pumping speed S = dV ⁄ dt is a volume flow, i.e. the volume of gas passing through the pump; commonly used by manufacturers to characterize their pumps.
Gas load q ~ dm ⁄ dt is equivalent to a mass flow, i.e. the total number of molecules passing through the pump.

Evacuation time

image
Table 1:Evacuation time as a function of pumping speed S and operating pressure p
S [m³ ⁄ h]10⁰ mbar10⁻¹ mbar10⁻² mbar10⁻³ mbar10⁻⁴ mbar
105 min6 min7 min9 min10 min
203 min3 min4 min5 min5 min
302 min2 min3 min3 min4 min
402 min2 min2 min3 min3 min
501 min2 min2 min2 min2 min
601 min1 min2 min2 min2 min
701 min1 min1 min2 min2 min
801 min1 min1 min2 min2 min
901 min1 min1 min1 min2 min
1001 min1 min1 min1 min1 min
For p0 = 1 013 mbar and V = 0.1 m3; rounded to the next integer

Effective pumping speed (S)

  • Pumping speed is not constant, but varies with pressure.
  • Different pumping speeds for different types of gases
  • Conductance C (how effectively the gas is transported in the system) may greatly influence the effective pumping speed:
image

Matching roughing pump and vacuum pump

Usually, the process defines gas load q and operating pressure p, and therefore the required effective pumping speed of the vacuum pump:
image
If a roughing pump is required, it has to handle the same gas load as the vacuum pump (because they are running in series).
image

Requirements for high vacuum

Table 2:Requirements for high vacuum
For high vacuum (and lower pressure) evacuation time and maintainable pressure are mainly determined by desorption and leakage rates.
≥ 10⁻⁴ mbarEvacuation time is the sum of time for roughing pump and for vacuum pump (pressure may even be achieved with roughing pump only)
~ minutes
< 10⁻⁴ mbarEvacuation is dominated by desorption fromchamber surfaceClean, smooth chamber surface is cruciale.g. turbo pump
< 10⁻⁶ mbarDesorption of sealing material becomes relevantTurbo pump (possibly with assistance of ion getter pump)
< 10⁻⁸ mbarMetal sealing is requiredIon getter pump is highly recommended
< 10⁻¹⁰ mbarLeakage and permeation through chamber walls are no longer negligible